Shaping the future of radiotherapy: the role of electron beams and flash techniques
- Authors: Parshenkov M.A.1, Skovorodko P.P.1, Petrusevich D.A.1, Makaeva S.N.1, Osipova S.G.1, Ibragimova G.R.1, Istyagina A.O.1, Balaeva K.A.1, Rodionova G.M.1
-
Affiliations:
- First Moscow State Medical University (Sechenov University)
- Issue: Vol 29, No 3 (2025): ONCOLOGY
- Pages: 283-301
- Section: ONCOLOGY
- URL: https://journals.rudn.ru/medicine/article/view/46084
- DOI: https://doi.org/10.22363/2313-0245-2025-29-3-283-301
- EDN: https://elibrary.ru/OGBXJP
- ID: 46084
Cite item
Full Text
Abstract
Relevance. Radiation therapy (RT) remains a cornerstone of oncology, offering targeted treatment for various cancers. With its roots tracing back to the discovery of X-rays by Wilhelm Röntgen and radium research by Marie Curie, RT has evolved into a sophisticated field encompassing a range of techniques. However, the rising global cancer burden highlights the need for continuous advancements to enhance efficacy while minimizing collateral damage. Traditional modalities such as X-rays and gamma rays have established their role in cancer treatment, yet they often lead to unintended damage to healthy tissues. Electron therapy has emerged as a promising alternative, leveraging distinct dosimetric properties that enable precise targeting with limited penetration depth. Low-energy electron beams are ideal for superficial tumors, while Very High-Energy Electrons (VHEEs) extend the reach to deep-seated tumors, rivalling proton and heavy-ion therapies. Furthermore, the FLASH effect — a phenomenon reducing healthy tissue toxicity at ultra-high dose rates, offers a breakthrough in electron therapy, improving patient quality of life. Despite these advancements, challenges persist. Limited penetration depth, secondary radiation from bremsstrahlung, and complexities in dose delivery systems constrain broader clinical adoption. Moreover, unresolved biological uncertainties, such as variability in relative biological effectiveness (RBE), necessitate further research. This review explores the historical evolution, unique benefits, and limitations of electron therapy compared to traditional modalities. It highlights advancements like VHEEs, FLASH therapy, and hybrid approaches, while addressing technological challenges and the future potential of electron beams in oncology. Conclusion. Integrated with recent technological breakthroughs, electron therapy may redefine the future of radiotherapy by offering safer, more precise, and individualized cancer treatment strategies.
Full Text
Introduction
Radiation therapy (RT), one of the fundamental methods for cancer treatment worldwide, has a rich history spanning over 130 years (Figure 1) [1]. Wilhelm Röntgen’s discovery of X-rays in 1895 marked a revolutionary turning point in medicine, enabling the use of ionising radiation to treat various types of malignant tumours. Building on this foundation, Marie Curie’s pioneering research on Radium [2, 3], for which she received a Nobel Prize, laid the groundwork for modern radiotherapy, making it a critical component in oncology and helping to save countless lives. Today, radiotherapy is supported by a multidisciplinary team of radiation oncologists, medical physicists, dosimetrists, and radiotherapists, all working collaboratively to ensure precise planning and delivery of treatment.
Fig. 1. Milestones in the evolution of Radiation Oncology: from the discovery of X-rays by Wilhelm Röntgen to modern advancements like FLASH and VHEE therapy
The field faces growing challenges as cancer cases are projected to increase by 77% by 2050, according to the World Health Organization [4]. This anticipated rise underscores the pressing need for continuous advancements in radiation therapy to meet the increasing demand for effective cancer treatment.
Radiation therapy has become an indispensable modality in modern oncology, serving both as a standalone treatment for malignant neoplasms (MN) and as an adjunct to other therapeutic approaches, including surgical interventions, chemotherapy, targeted molecular therapies, and immunotherapy [5]. The choice of radiation therapy method depends on various factors, such as the type, localization, and stage of cancer, highlighting the necessity of personalized treatment strategies for each patient.
Among its numerous applications, radiation therapy has shown particular efficacy in the treatment of breast cancer, where it plays a central role in both curative and palliative settings [6]. Techniques such as three-dimensional conformal radiation therapy (3D CRT), intensity-modulated radiation therapy (IMRT), and volumetric-modulated arc therapy (VMAT) are widely employed [7–9]. Their effectiveness varies depending on tumor location and stage, with VMAT often preferred for left-sided breast cancer to minimize cardiac and pulmonary toxicity, while 3D CRT remains the method of choice for right-sided cases to reduce myocardial exposure [10]. Additionally, hypofractionation schedules, which reduce the overall treatment duration, have been increasingly adopted to improve patient convenience while maintaining therapeutic efficacy [11].
Radiation delivery methods are broadly classified into external beam radiation therapy (EBRT), brachytherapy, and systemic radiation. EBRT utilizes high-energy beams such as X-rays, gamma rays, or particles like electrons and protons, targeting tumors with high precision through advanced imaging and planning technologies [12]. On the other hand, brachytherapy involves placing radioactive sources directly within or adjacent to the tumor, allowing for localized dose escalation while minimizing exposure to surrounding healthy tissues [13]. These approaches, combined with emerging techniques such as adaptive radiation therapy and proton therapy, aim to enhance the therapeutic index by achieving higher tumor control rates with reduced toxicity to normal tissues (Figure 2).
Fig. 2. Overview of modern Radiation Therapy Techniques and their applications
Recent advancements in radiation oncology have not only refined traditional approaches but have also integrated multi-modality treatment strategies. For instance, the combination of different radiation techniques with systemic therapies has enabled the targeting of radioresistant tumors, increasing the likelihood of achieving durable local control [14]. Furthermore, advanced imaging and treatment planning systems have improved dose conformity, enabling clinicians to optimize radiation delivery while protecting critical structures and minimizing long-term side effects [15].
However, despite the crucial position that RT occupies in the treatment of multiple cancers, the effects of ionizing radiation on the body are complex and affect biological structures at the cellular, molecular and other levels. Conventional methods of irradiation, including X-ray and gamma therapy, although effective, often result in collateral damage to healthy tissue, highlighting the need to develop better approaches that maintain therapeutic efficacy while minimizing harm.
Electron therapy has emerged as a promising alternative in this regard, offering distinct physical properties that allow for more targeted radiation delivery with reduced penetration depth. This feature enables electron therapy to be highly effective for treating tumors located near the skin surface or in shallow tissue areas [16], making it less likely to impact deeper organs and structures. Unlike photon-based radiotherapy, which can penetrate deeply and affect both the tumor and surrounding tissue, electron therapy’s limited depth of action presents significant potential for reducing treatment-related side effects [17, 18].
Recent advances have broadened the potential of electron therapy. Very High-Energy Electrons (VHEEs, 100–250 MeV) now show promise for treating deep-seated tumors, approaching the precision of proton and carbon ion therapies while offering a more accessible and cost-effective alternative [19]. A key advantage of electron beams is their compatibility with the FLASH effect — a phenomenon in which ultra-high dose rates reduce toxicity to healthy tissue without sacrificing tumor control [20]. By harnessing FLASH, electron therapy may dramatically reduce side effects and improve patients’ quality of life.
We explore the potential of electron therapy within the broader scope of radiotherapy, emphasizing its unique benefits, limitations, and its comparison with other radiation types, such as X-rays and gamma rays. This paper also addresses current technological challenges and examines the promise of advanced techniques, such as the FLASH effect and Very High-Energy Electrons, in establishing electron therapy as a standalone and safer modality in radiation treatment. By offering a comprehensive overview of its advantages, limitations, and areas ripe for further research, this review aims to support advancements in oncology to enhance radiotherapy outcomes for patients.
Fundamental principles of radiation therapy
Radiation therapy is a cornerstone of oncology, utilizing high-energy particles or waves [21]. The targeted damage induces cell death in malignant tissues, making radiation therapy an effective strategy for treating a wide range of cancers. The success of this approach depends on several factors, including the type of radiation, linear energy transfer (LET), total dose, fractionation schedules, and the radiosensitivity of tumor cells [22].
A critical determinant of the biological effectiveness of radiation therapy is linear energy transfer (LET), which measures the energy deposited by radiation per unit distance travelled in tissue [23]. The High-LET radiation, such as alpha particles or neutrons, creates dense ionization tracks, resulting in more severe macromolecular damage and increased cell lethality compared to low-LET radiation like X-rays or gamma rays [24]. This heightened damage includes double-strand breaks (DSBs) in DNA [25], a pivotal factor in inducing apoptosis, necrosis, or cellular senescence. Also, The High-LET radiation could engage specific DNA repair pathways, with non-homologous end-joining (NHEJ) being particularly critical [24]. Notably, defects in NHEJ pathways lead to heightened radiosensitivity across various LET ranges, emphasizing the central role of repair mechanisms in determining radiation-induced cell fate (Figure 3) [25, 26].
Fig. 3. Comparison of DNA damage and repair mechanisms induced by low-LET and high-LET radiation
The relationship between LET and relative biological effectiveness (RBE) is intricate. While the high-LET radiation is generally associated with increased RBE, this relationship varies depending on radiation quality, cellular characteristics, and biological endpoints under investigation. Recent studies highlight inconsistencies in how LET is defined and averaged across the literature, potentially undermining RBE modelling and its clinical translation [27]. Resolving these inconsistencies is paramount to enhancing the precision of treatment planning and improving therapeutic outcomes.
Beyond DNA damage, the biological effects of LET encompass secondary mechanisms such as oxidative stress and bystander effects. Oxidative stress, driven by reactive oxygen species (ROS), exacerbates cellular damage, while bystander signalling propagates radiation effects to neighbouring cells, influencing both tumor control and normal tissue toxicity (Fig. 4) [28–30]. Understanding these mechanisms is particularly critical for advancing novel radiation modalities, including proton therapy and heavy-ion radiation, which inherently rely on the interplay between LET and biological response.
Fig. 4. Mechanisms of Radiation-induced cellular and Bystander effects
By integrating these insights, radiation therapy continues to advance as a precise and adaptable treatment modality in oncology. Nevertheless, challenges persist, including the variability of RBE predictions and the absence of universally accepted LET parameters. These hurdles underscore the need for ongoing research to refine biological models and optimize clinical application, paving the way for more effective and individualized cancer treatments.
Electron Irradiation
Electron irradiation presents significant potential in radiation oncology due to its distinct physical properties, particularly penetration depth and scattering patterns [31]. These properties are critical for optimizing electron-based treatments, especially in scenarios where precision is required to target tumors while preserving adjacent healthy tissues.
The penetration depth of electron beams is highly dependent on their energy and the atomic composition of the irradiated material [32]. Low-energy electron beams (6–20 MeV) are well-suited for treating superficial malignancies, as their limited penetration depth ensures dose delivery primarily to the skin and shallow tissues, sparing underlying structures [33]. On the other hand, Very High-Energy Electrons (VHEEs), with energies exceeding 100 MeV, provide the capability to treat tumors located at depths of 5–15 cm while maintaining dose conformity even in heterogeneous tissues such as the lungs and bones [34, 35]. These properties position VHEEs as a potential alternative to photon and proton therapies for certain clinical applications (Figure 5).
Fig. 5. Theoretical depth-dose distribution of radiation in water for various particle beams. The graph illustrates the absorbed dose of radiation as a function of penetration depth for different particle types, including 6 MeV photons, 15 MeV electrons, 150 MeV protons, and 250 MeV Very High-Energy Electrons (VHEEs)
Recent innovations, such as FLASH radiotherapy, have further expanded the potential of electron beams. FLASH therapy involves delivering ultra-high dose rates (>40 Gy/s), which have been shown to reduce normal tissue toxicity while preserving tumor control [36]. Although most FLASH studies in medical literature have focused on low-energy electrons, VHEEs could extend this modality to deeper-seated tumors. The precise mechanisms underlying the FLASH effect remain under investigation, with current hypotheses suggesting reduced oxygenation and suppression of reactive oxygen species as key factors [37].
Electron scattering is central to treatment planning, as elastic, inelastic, and bremsstrahlung interactions shape dose distribution, especially in anatomically complex regions [38, 39]. Monte Carlo simulations remain the gold standard for modeling these effects, accurately predicting energy deposition, range straggling, and lateral spread across media [40]. Notably, VHEE beams exhibit a stable, narrow profile at depth, delivering more uniform doses than photons in heterogeneous tissues [41]. Fermi-Eyges models, closely aligned with Monte Carlo data, further enable faster, high-precision calculations for VHEE therapy [42].
Advances in beam delivery systems, such as active scanning techniques and pencil beam scanning, have significantly enhanced the precision of electron therapy [43]. These technologies enable highly focused dose delivery, reducing exposure to surrounding tissues and accommodating complex tumor geometries. For example, a 160 MeV laser-wakefield accelerated electron beam has been focused to a 2.3 × 2.6 mm2 spot, demonstrating the feasibility of stereotactic radiotherapy with sub-millimeter precision [41, 43]. Furthermore, integrating Monte Carlo methods with imaging modalities like dual-energy CT has refined treatment planning, improving the accuracy of stopping power ratios and dose predictions [44].
Combining different electron beam energies offers additional flexibility in treatment. For instance, mixed-energy beams can effectively expand the treatment region while maintaining steep dose “fall-offs”, enabling the precise targeting of irregularly shaped tumors [45]. This versatility highlights the adaptability of electron therapy for a wide range of clinical scenarios and emphasizes the different situation of use.
Gamma Irradiation
Another approach we would like to focus on is gamma irradiation. Gamma radiation, a high-energy form of electromagnetic radiation, stands out for its exceptional penetration capabilities, making it indispensable in radiotherapy and other applications [46]. Its uncharged, high-energy photons enable gamma rays to deeply penetrate matter, interact with tissues via ionization, and cause excitation at the atomic level [47]. These interactions underlie gamma radiation’s ability to inactivate pathogens, extend food shelf-life, and perform critical roles in medical and industrial applications. In radiotherapy, gamma rays interact primarily through the Compton effect, leading to ionization and subsequent biological effects, including DNA damage [48].
Typically, gamma rays are generated through radioactive decay, nuclear fission, or advanced particle accelerators [49]. Cobalt60 and Cesium137 are among the most widely used isotopes, offering reliable gamma ray emission for therapeutic and research purposes [50]. These rays exhibit wavelengths ranging from approximately 0.25 to 0.005 angstroms, with energy levels spanning tens of thousands to several million electronvolts [51]. Due to their high energy, effective shielding using dense materials like lead or concrete is essential in clinical and, for example, industrial settings.
Additionally, gamma radiation is integral to radio-surgical techniques, with the Gamma Knife exemplifying its precision and effectiveness in treating localized conditions (Figure 6) [52]. This non-invasive method uses 201 Cobalt60 sources to deliver focused beams, achieving high doses at targeted lesions while sparing adjacent healthy tissue [53]. The Gamma Knife has demonstrated success in managing arteriovenous malformations (AVMs) and inoperable brain tumors, with studies linking biologically effective doses to higher obliteration rates [54]. Recent advancements, such as frameless Gamma Knife radiosurgery, have enhanced patient experience, enabling fractionation without compromising spatial resolution [55]. These developments underscore gamma radiation’s role in advancing neurosurgical outcomes.
Fig. 6. Schematic representation of Gamma Knife radiosurgery. The diagram illustrates the precise targeting of a brain tumor (target tumour) using converging gamma rays, delivered through a specialized helmet. The focused radiation beams spare the surrounding healthy brain tissue, demonstrating the non-invasive and localized nature of the treatment
Despite its extensive applications, the effective and safe use of gamma radiation relies on accurate dosimetry and individualized treatment planning [56]. Dosimetric verification methods, including gamma analysis, are vital for ensuring precise dose delivery in treatments like Gamma Knife radiosurgery. However, standard activity prescriptions often fail to account for individual patient differences, necessitating advancements in personalized dosimetry techniques. Challenges such as non-standardized protocols and the complexity of dose distribution modelling require further research and technical innovation to optimize patient outcomes.
Recent innovations, such as “mini-beam” collimator designs, have improved gamma-based dose conformity, allowing for narrower beams with high peak-to-valley dose ratios. Techniques like volumetric modulated arc therapy (VMAT) and intensity-modulated radiation therapy (IMRT) further enhance precision by modulating collimator positioning, dose rates, and gantry speeds [57]. These advancements minimize radiation exposure to surrounding tissues, reducing side effects and enhancing treatment outcomes.
Finally, it is important to consider that gamma radiation’s biological efficacy is enhanced by its linear energy transfer, especially in radiosurgical applications. While low-LET radiation like gamma rays primarily causes single-strand DNA breaks, high-LET radiation, including α-particles and heavy ions, induces complex DNA damage such as clustered lesions and multiple double-strand breaks (DSBs) [58]. These DSBs activate homologous recombination repair pathways more effectively than non-homologous end-joining, contributing to the higher biological effectiveness of high-LET radiation [59]. Leveraging these differences in DNA repair mechanisms can optimize cancer therapy and aid in developing targeted radiosensitizers.
X-Ray Irradiation
X-rays, a high-energy form of electromagnetic radiation, have revolutionized both medical and industrial applications due to their remarkable penetration capabilities (Figure 7) [60]. These waves, with wavelengths ranging from 10 pm to 10 nm and energies typically between 100 eV and 500 keV, possess unique properties that allow them to penetrate dense materials, enabling imaging and therapeutic interventions across a broad spectrum of fields (Figure 8) [61]. In the medical realm, X-rays are indispensable for diagnostics, including radiography and computed tomography (CT), and for therapeutic applications such as sterilization of single-use devices, where X-ray sterilization is emerging as an alternative to traditional gamma radiation [62].
X-ray imaging remains a cornerstone of clinical diagnostics, with techniques like radiography and CT relying on the differential absorption of X-rays by various tissues to produce high-resolution internal images. These modalities are critical for detecting cancers, bone fractures, and other medical conditions [63]. Recent advances include the development of dual-energy and multi-spectral CT, which enable quantitative tissue analysis and material characterization, offering improved diagnostic accuracy and interventional guidance [64]. However, the associated risks of radiation exposure necessitate strategies to minimize harm to both patients and medical staff.
The mechanism of X-ray generation, known as bremsstrahlung or “braking radiation”, occurs when high-energy electrons decelerate as they interact with atomic nuclei. This process produces a broad spectrum of X-ray energies, making it the cornerstone of X-ray production in medical and industrial settings [61, 65]. Recent advancements in bremsstrahlung include the development of electron wavefunction shaping techniques, which have demonstrated enhanced X-ray intensity and emission control when electrons interact periodically with crystalline materials [66]. Additionally, accurate modelling of bremsstrahlung production has refined predictions of X-ray energy and angular distributions, enhancing applications ranging from industrial inspection to advanced imaging techniques [67].
The biological effects of X-rays are primarily mediated through their interaction with DNA, leading to complex DNA damage (CDD), including double-strand breaks (DSBs) [68]. These lesions activate the DNA damage response (DDR), initiating pathways such as cell cycle arrest, DNA repair, and apoptosis [69]. Radiotherapy leverages this property to target cancer cells, exploiting their impaired repair mechanisms to induce cell death. Recent research highlights the potential of targeting DNA repair pathways to enhance radiosensitivity in tumors, thereby overcoming radioresistance while protecting normal tissue [70]. Emerging approaches, such as mitophagy enhancement, further illustrate innovative strategies to amplify radiation-induced DNA damage, paving the way for improved therapeutic outcomes [71].
Fig. 7. Comparative penetrating power of various radiation types through different materials
Fig. 8. Representation of the electromagnetic spectrum highlighting X-rays and other radiation types by wavelength range
X-rays also enable hybrid imaging techniques, such as X-ray-induced optical molecular sensing. These methods combine scintillation or Cherenkov radiation with molecular probes to achieve high-resolution imaging of tissue properties like oxygenation and pH, offering novel diagnostic opportunities [72]. Advances in scintillation materials, such as all-inorganic perovskite nanocrystals, have further improved X-ray detection efficiency, enhancing both medical and industrial imaging [62].
In industrial applications, X-rays are widely used for non-destructive testing and cargo screening. Dual-angle X-ray systems, such as those designed with 9 MV linear accelerators, have demonstrated improved contrast sensitivity and spatial resolution, ensuring precise inspections [73]. The development of wearable dosimeters for personnel protection, incorporating artificial intelligence and Internet of Things technologies, reflects the ongoing innovations aimed at optimizing X-ray safety in various fields.
Despite these advancements, challenges persist. Secondary radiation, such as that generated by bremsstrahlung, and the need for precise beam collimation complicate X-ray applications in both medical and industrial contexts [74]. Additionally, optimizing dose delivery while minimizing radiation-induced harm remains a critical area for research and development. As our understanding of X-ray interactions with biological tissues and materials deepens, future innovations will undoubtedly expand the potential of X-rays as a versatile tool for imaging, therapy, and diagnostics.
In-depth analysis of electron therapy in radiation treatment
To begin with, any therapeutic technique (whether established or emerging) must undergo rigorous evaluation from multiple perspectives, in context both its current advantages and inherent limitations. This approach is especially crucial in radiotherapy, where therapeutic decisions directly impact patient health and survival outcomes. Electron therapy, with its potential for high precision and adaptability, presents significant benefits across various clinical applications. However, a thorough understanding of its role demands not only an appreciation of its strengths but also a critical examination of its limitations and associated challenges. In this section, we aim to provide a comprehensive overview of electron therapy as an advancing modality. Through an analysis of current literature and recent innovations, we seek to offer a balanced perspective on its clinical promise, as well as the barriers that may influence its broader adoption in oncology.
Advantages of electron therapy in radiation oncology
Unique dosimetric properties of electron therapy
Electron therapy offers a distinct advantage in radiation oncology due to its unique dose distribution characteristics. Unlike traditional photon radiation, electron beams provide a homogeneous dose profile within the target area, with a sharp dose fall-off beyond a specific depth. This property makes them particularly effective for superficial tumors and cutaneous lymphomas, minimizing damage to underlying healthy tissues [75]. The steep dose gradients of electron beam also allow precise treatment of irregularly shaped targets while sparing adjacent organs and tissues [31]. Very High-Energy Electron (VHEE) beams, operating at energies above 100 MeV, have expanded the scope of electron therapy to include deep-seated tumors up to 15 cm depth. These beams offer improved dose conformity and reduced lateral scattering compared to photon beams, making them suitable for treating challenging anatomical regions [76].
Adaptability for Targeting Tumorsat Various Depths
One of the most notable benefits of electron therapy is its adaptability. The ability to adjust beam energy allows for tailored treatments targeting tumors at varying depths. Low-energy beams are ideal for shallow lesions, while higher-energy beams address deeper tumors. For instance, electron therapy has shown efficacy in breast cancer treatments, particularly as a boost modality after breast-conserving surgery. Studies demonstrate reduced radiation doses to critical structures such as the heart and contralateral breast, maintaining high target coverage [77]. Recent advancements in hybrid techniques, such as combining electron and photon therapy, have further optimized treatment outcomes [17]. The use of 3D-printed modulated electron boluses (MEBs) has improved dose conformity and reduced exposure to organs at risk, including the ipsilateral lung and left anterior descending artery [78]. These innovations underline the potential of electron therapy to enhance patient quality of life.
Advancements in Very High-Energy Electron Therapy
VHEE therapy becomes a pioneering technique in modern radiotherapy due to its unique physical properties and potential to treat complex cancer cases [19]. Using electrons with energies in a wide range, it is characterised by its ability to deliver therapeutic doses with remarkable precision, which makes it particularly suitable for the treatment of tumours in heterogeneous and anatomically complex areas such as the thoracic and pelvic cavities (including the retroperitoneum and pelvic organs) [34].
Comparative analyses have demonstrated that VHEE beams achieve dosimetric performance similar to advanced modalities like proton therapy, offering an effective yet potentially more accessible alternative for institutions without proton beam facilities. Advanced beam delivery systems, including collimation and modulation techniques, have further enhanced this precision. Recent innovations in treatment planning, such as adaptive optimization algorithms, have enabled tailored dose distributions to accommodate tumors with irregular geometries or varying densities [79], underscoring the versatility of VHEE in clinical practice. As research progresses, the application of VHEE beams continues to evolve, with ongoing studies focusing on optimizing treatment protocols, refining beam delivery systems, and enhancing patient outcomes. By addressing existing technical challenges and expanding its clinical capabilities, VHEE therapy holds immense promise as a cornerstone of next-generation radiation oncology.
Innovative Delivery Systems and Emerging Technologies
Technological innovations continue to expand the versatility of electron therapy. Laser Wakefield Acceleration (LWFA) offers a compact and cost-effective method for generating high-energy electron beams, potentially making electron therapy more accessible in clinical settings [80]. Researchers have demonstrated the generation of ultrarelativistic electron beams up to 50 MeV using terawatt-scale lasers, paving the way for advanced radiotherapy devices [81]. Advanced collimation and scanning systems are also under development to support the precise delivery of VHEE beams. These systems incorporate optimized magnetic fields and high-Z foils to sustain ultra-high dose rates while ensuring dose uniformity. Combined with PBS techniques, these innovations enhance the precision and safety of electron therapy, reinforcing its role as a cornerstone in modern radiation oncology.
Biological Applications and Potential of Electrons
Electron therapy has emerged as a versatile modality in radiation oncology, offering unique biological and dosimetric advantages. Low-energy electrons (LEEs) play a crucial role in inducing DNA damage, primarily through ionization and excitation, leading to double-strand breaks (DSBs) and oxidative stress in tumor cells [82]. These effects are highly localized, minimizing collateral damage to healthy tissues. Advanced computational tools, such as Monte Carlo simulations, have provided insights into the interactions of LEEs with biological molecules, aiding in precise future treatment planning.
Recent innovations, such as Very High-Energy Electrons have expanded the potential of electron therapy to treat deep-seated tumors [32]. Studies demonstrate that VHEEs can achieve dosimetric quality comparable to proton therapy, particularly for challenging cancers such as glioblastoma and prostate cancer [57]. Furthermore, the integration of electron therapy with emerging modalities like FLASH radiotherapy has shown promise in enhancing the therapeutic index. VHEE beams are particularly suited for FLASH applications, offering both biological and ballistic advantages for treating deep-seated tumors.
Limitations and challenges of electron therapy
Despite its significant potential, electron therapy faces several limitations and challenges that restrict its broader clinical application. One of the primary limitations is the restricted penetration depth and dose modulation of conventional electron beams. Their shallow penetration depth makes them less effective for treating deep-seated tumors, even though there’s a lot of research of this flaw. Thus, Very High-Energy Electrons partially address this limitation by providing improved dose matching for tumors located at a depth of 5–15 cm. However, achieving such precision requires energies above 100 MeV, for greater penetration depths — up to 30 or more centimeters deep, and advanced beam modulation techniques and highly specialized delivery systems. These setups, while promising, are not yet widely available and require further technological development.
Another critical issue is the presence of secondary radiation components, particularly Bremsstrahlung photons generated by high-energy electron beams. These secondary photons can contribute up to 8% of the absorbed dose [83], increasing the risk of unintended biological effects, particularly in high-energy applications. This underscores the importance of careful machine design and optimization of beam delivery systems to mitigate these risks. Accurate dosimetry and beam monitoring also present significant challenges, especially in ultra-high dose-rate applications such as FLASH therapy. Existing detectors struggle with saturation effects, limiting their ability to measure doses accurately at the speeds required for FLASH [84]. The development of robust monitoring tools and the standardization of dosimetry protocols are essential to ensure precise and safe dose delivery.
The scattering dynamics and lateral penumbra of electron beams further complicate their use, particularly in heterogeneous tissues. Advanced techniques such as pencil beam scanning (PBS) and magnetic focusing have demonstrated potential in reducing lateral penumbra and enhancing dose conformity [85]. However, these technologies are still in experimental stages and require further refinement to become clinically viable. In addition, the technical challenges associated with FLASH and VHEE therapy remain a significant bottleneck. Achieving the ultra-high dose rates necessary for FLASH therapy involves overcoming hurdles such as rapid energy adjustments and efficient pencil beam scanning for large treatment fields [84]. Current systems lack the capability to deliver FLASH dose rates across extensive areas, limiting their practical application. The development of advanced technologies is necessary to address these issues.
Finally, biological uncertainties surrounding electron therapy must be resolved. The relative biological effectiveness (RBE) of electrons, often assumed to be unity, may vary depending on factors such as energy levels, dose rates, and tissue types. While preliminary studies have reported differences in DNA double-strand break (DSB) induction between conventional and VHEE beams [86], the clinical significance of these findings remains unclear. A deeper understanding of the underlying radiobiological mechanisms is crucial to optimizing treatment outcomes and minimizing potential side effects.
Conclusion
Electron therapy represents a promising advancement in radiation oncology, offering unique physical and biological properties that enable precise and adaptable treatment strategies. Its dosimetric advantages, particularly the ability to deliver highly localized doses with minimal collateral damage, position it as a valuable modality for both superficial and deep-seated tumors. Recent innovations, such as Very High-Energy Electrons and FLASH radiotherapy, have expanded the scope of electron therapy, demonstrating significant potential for improving therapeutic outcomes and patient quality of life.
But several challenges remain. The limited availability of advanced delivery systems, complexities in beam modulation, and biological uncertainties, like variations in relative biological effectiveness, underscore the need for continued research and technological development. Addressing these limitations through interdisciplinary collaboration and innovation will be essential for unlocking the full potential of electron therapy as a standard clinical modality.
By bridging the gap between emerging technologies and clinical application, future efforts in electron therapy hold the promise of transforming radiotherapy practices, enhancing precision, and ultimately improving outcomes for cancer patients worldwide.
About the authors
Mikhail A. Parshenkov
First Moscow State Medical University (Sechenov University)
Author for correspondence.
Email: misjakj@gmail.com
ORCID iD: 0009-0004-7170-8783
SPIN-code: 7012-6284
Moscow, Russian Federation
Polina P. Skovorodko
First Moscow State Medical University (Sechenov University)
Email: misjakj@gmail.com
ORCID iD: 0009-0000-5624-4731
Moscow, Russian Federation
Danila A. Petrusevich
First Moscow State Medical University (Sechenov University)
Email: misjakj@gmail.com
ORCID iD: 0009-0005-2283-6372
Moscow, Russian Federation
Shagun N. Makaeva
First Moscow State Medical University (Sechenov University)
Email: misjakj@gmail.com
ORCID iD: 0009-0008-3396-6684
Moscow, Russian Federation
Svetlana G. Osipova
First Moscow State Medical University (Sechenov University)
Email: misjakj@gmail.com
ORCID iD: 0009-0006-6457-9543
Moscow, Russian Federation
Gumru R. Ibragimova
First Moscow State Medical University (Sechenov University)
Email: misjakj@gmail.com
ORCID iD: 0009-0007-0478-7137
Moscow, Russian Federation
Alexandra O. Istyagina
First Moscow State Medical University (Sechenov University)
Email: misjakj@gmail.com
ORCID iD: 0009-0001-3295-8462
Moscow, Russian Federation
Kristina A. Balaeva
First Moscow State Medical University (Sechenov University)
Email: misjakj@gmail.com
ORCID iD: 0009-0002-2930-6006
Moscow, Russian Federation
Galina M. Rodionova
First Moscow State Medical University (Sechenov University)
Email: misjakj@gmail.com
ORCID iD: 0000-0002-0536-9590
SPIN-code: 5657-9984
Moscow, Russian Federation
References
- Trofimova OP, Tkachev SI, Yurieva TV. The Past and Present of Radiation Therapy in Oncology. Clinical Oncohematology. Fundamental Research and Clinical Practice. 2013;6(4):355–364. (In Russian).
- Brailsford JF. Roentgen’s discovery of X-rays; their application to medicine and surgery. British Journal of Radiology. 1946;19(227):453–461. doi: 10.1259/0007-1285-19-227-453
- Abergel R, Aris J, Bolch WE, Dewji SA, Golden A, Hooper DA, Margot D, Menker CG, Paunesku T, Schaue D, Woloschak GE. The enduring legacy of Marie Curie: impacts of radium in 21st century radiological and medical sciences. International Journal of Radiation Biology. 2022;98(3):267–275. doi: 10.1080/09553002.2022.2027542
- World Health Organization. Global cancer burden growing amidst mounting need for services. 2024 Feb 1 [Accessed Aug, 62024]. Available from: https://www.who.int/news/item/01–02–2024‑global-cancer-burden-growing–amidst-mounting-need-for-services
- Upadhyay R, Bazan JG. Advances in Radiotherapy for Breast Cancer. Surgical Oncology Clinics of North America. 2023;32(3):515–536. doi: 10.1016/j.soc.2023.03.002
- Meattini I, Livi L, Lorito N, Becherini C, Bacci M, Visani L, Fozza A, Belgioia L, Loi M, Mangoni M, Lambertini M, Morandi A. Integrating radiation therapy with targeted treatments for breast cancer: From bench to bedside. Cancer Treatment Reviews. 2022;108:102417. doi: 10.1016/j.ctrv.2022.102417
- Ambrose L, Stanton C, Lewis L, Lamoury G, Morgia M, Carroll S, Bromley R, Atyeo J. Potential gains: Comparison of a mono-isocentric three-dimensional conformal radiotherapy (3D-CRT) planning technique to hybrid intensity-modulated radiotherapy (hIMRT) to the whole breast and supraclavicular fossa (SCF) region. Journal of Medical Radiation Sciences. 2022;69(1):75–84. doi: 10.1002/jmrs.473
- Jagsi R, Griffith KA, Moran JM, Matuszak MM, Marsh R, Grubb M, Abu-Isa E, Dilworth JT, Dominello MM, Heimburger D, Lack D, Walker EM, Hayman JA, Vicini F, Pierce LJ; Michigan Radiation Oncology Quality Consortium. Comparative Effectiveness Analysis of 3D-Conformal Radiation Therapy Versus Intensity Modulated Radiation Therapy (IMRT) in a Prospective Multicenter Cohort of Patients With Breast Cancer. International Journal of Radiation Oncology, Biology, Physics. 2022;112(3):643–653. doi: 10.1016/j.ijrobp.2021.09.053
- Kamer S, Yilmaz Susluer S, Balci Okcanoglu T, Kayabasi C, Ozmen Yelken B, Hoca S, Tavlayan E, Olacak N, Anacak Y, Olukman M, Gunduz C. Evaluation of the effect of intensity-modulated radiotherapy (IMRT) and volumetric-modulated arc radiotherapy (VMAT) techniques on survival response in cell lines with a new radiobiological modeling. Cancer Medicine. 2023;12(19):19874–19888. doi: 10.1002/cam4.6593
- Tumanova K, Nidal S, Popodko A, Stolbovoy A. Most Appropriate Radiation Therapy Techniques for the Breast Cancer Treatment: Dosimetric Analysis of Three Different Radiation Therapy Methods. Radiotherapy and Clinical Oncology. 2023;26:74–97. doi: 10.31487/j.rco.2023.01.01
- Yan M, Gouveia AG, Cury FL, Moideen N, Bratti VF, Patrocinio H, Berlin A, Mendez LC, Moraes FY. Practical considerations for prostate hypofractionation in the developing world. Nature Reviews Urology. 2021;18(11):669–685. doi: 10.1038/s41585–021–00498–6
- Apisarnthanarax S, Barry A, Cao M, Czito B, DeMatteo R, Drinane M, Hallemeier CL, Koay EJ, Lasley F, Meyer J, Owen D, Pursley J, Schaub SK, Smith G, Venepalli NK, Zibari G, Cardenes H. External Beam Radiation Therapy for Primary Liver Cancers: An ASTRO Clinical Practice Guideline. Practical Radiation Oncology. 2022;12(1):28–51. doi: 10.1016/j.prro.2021.09.004
- Neugebauer J, Blum P, Keiler A, Süß M, Neubauer M, Moser L, Dammerer D. Brachytherapy in the Treatment of Soft-Tissue Sarcomas of the Extremities — A Current Concept and Systematic Review of the Literature. Cancers (Basel). 2023;15(4):1133. doi: 10.3390/cancers15041133
- Peng X, Wei Z, Gerweck LE. Making radiation therapy more effective in the era of precision medicine. Precision Clinical Medicine. 2020;3(4):272–283. doi: 10.1093/pcmedi/pbaa038
- Goodburn RJ, Philippens MEP, Lefebvre TL, Khalifa A, Bruijnen T, Freedman JN, Waddington DEJ, Younus E, Aliotta E, Meliadò G, Stanescu T, Bano W, Fatemi-Ardekani A, Wetscherek A, Oelfke U, van den Berg N, Mason RP, van Houdt PJ, Balter JM, Gurney-Champion OJ. The future of MRI in radiation therapy: Challenges and opportunities for the MR community. Magnetic Resonance in Medicine. 2022;88(6):2592–2608. doi: 10.1002/mrm.29450
- Vyas V, Palmer L, Mudge R, Jiang R, Fleck A, Schaly B, Osei E, Charland P. On bolus for megavoltage photon and electron radiation therapy. Medical Dosimetry. 2013;38(3):268–273. doi: 10.1016/j.meddos.2013.02.007
- Ma CM, Ding M, Li JS, Lee MC, Pawlicki T, Deng J. A comparative dosimetric study on tangential photon beams, intensity-modulated radiation therapy (IMRT) and modulated electron radiotherapy (MERT) for breast cancer treatment. Physics in Medicine and Biology. 2003;48(7):909–924. doi: 10.1088/0031–9155/48/7/308
- Van Eeden L, Sachse KN, Du Plessis FCP. Practical Dosimetry Considerations for Small MLC-Shaped Electron Fields at 60 cm SSD. Journal of Biomedical Physics and Engineering. 2022;12(1):101–108. doi: 10.31661/jbpe.v0i0.2004–1097
- Ronga MG, Cavallone M, Patriarca A, Leite AM, Loap P, Favaudon V, Créhange G, De Marzi L. Back to the Future: Very High-Energy Electrons (VHEEs) and Their Potential Application in Radiation Therapy. Cancers (Basel). 2021;13(19):4942. doi: 10.3390/cancers13194942
- Schüler E, Acharya M, Montay-Gruel P, Loo BW Jr, Vozenin MC, Maxim PG. Ultra-high dose rate electron beams and the FLASH effect: From preclinical evidence to a new radiotherapy paradigm. Medical Physics. 2022;49(3):2082–2095. doi: 10.1002/mp.15442
- Carlos-Reyes A, Muñiz-Lino MA, Romero-Garcia S, López-Camarillo C, Hernández-de la Cruz ON. Biological Adaptations of Tumor Cells to Radiation Therapy. Frontiers in Oncology. 2021;11:718636. doi: 10.3389/fonc.2021.718636
- Russ E, Davis CM, Slaven JE, Bradfield DT, Selwyn RG, Day RM. Comparison of the Medical Uses and Cellular Effects of High and Low Linear Energy Transfer Radiation. Toxics. 2022;10(10):628. doi: 10.3390/toxics10100628
- Kalholm F, Grzanka L, Traneus E, Bassler N. A systematic review on the usage of averaged LET in radiation biology for particle therapy. Radiotherapy and Oncology. 2021;161:211–221. doi: 10.1016/j.radonc.2021.04.007
- Roobol SJ, van den Bent I, van Cappellen WA, Abraham TE, Paul MW, Kanaar R, Houtsmuller AB, van Gent DC, Essers J. Comparison of High- and Low-LET Radiation-Induced DNA Double-Strand Break Processing in Living Cells. International Journal of Molecular Sciences. 2020;21(18):6602. doi: 10.3390/ijms21186602
- Li CY. Non-canonical roles of apoptotic and DNA double-strand break repair factors in mediating cellular response to ionizing radiation. International Journal of Radiation Biology. 2023;99(6):915–924. doi: 10.1080/09553002.2021.1948139
- Zhao B, Rothenberg E, Ramsden DA, Lieber MR. The molecular basis and disease relevance of non-homologous DNA end joining. Nature Reviews Molecular Cell Biology. 2020;21(12):765–781. doi: 10.1038/s41580-020-00297-8
- Smith EAK, Winterhalter C, Underwood TSA, Aitkenhead AH, Richardson JC, Merchant MJ, Kirkby NF, Kirby KJ, Mackay RI. A Monte Carlo study of different LET definitions and calculation parameters for proton beam therapy. Biomedical Physics & Engineering Express. 2021;8(1):10.1088/2057–1976/ac3f50. doi: 10.1088/2057–1976/ac3f50
- Nikitaki Z, Velalopoulou A, Zanni V, Tremi I, Havaki S, Kokkoris M, Gorgoulis VG, Koumenis C, Georgakilas AG. Key biological mechanisms involved in high-LET radiation therapies with a focus on DNA damage and repair. Expert Reviews in Molecular Medicine. 2022;24: e15. doi: 10.1017/erm.2022.6
- Sergey Koryakin, Kirill Petrushin, Mikhail Parshenkov, Zhanna Uruskhanova, Anastasiia Shchitkova, Elizabeth Pechnikova, Grigory Demyashkin. Kidney morphofunctional features after ascorbic acid administration in a model of acute radiation nephropathy. RUDN Journal of Medicine. 2024;28(3):301–310. doi: 10.22363/2313-0245-2024-28-3-37358. (In Russian)
- Buonanno M, Gonon G, Pandey BN, Azzam EI. The intercellular communications mediating radiation-induced bystander effects and their relevance to environmental, occupational, and therapeutic exposures. International Journal of Radiation Biology. 2023;99(6):964–982. doi: 10.1080/09553002.2022.2078006
- Svendsen K, Guénot D, Svensson JB, Petersson K, Persson A, Lundh O. A focused very high energy electron beam for fractionated stereotactic radiotherapy. Scientific Reports. 2021;11(1):5844. doi: 10.1038/s41598-021-85451-8
- Ghorbani M, Tabatabaei ZS, Vejdani Noghreiyan A, Vosoughi H, Knaup C. Effect of tissue composition on dose distribution in electron beam radiotherapy. Journal of Biomedical Physics and Engineering. 2015;5(1):15–24.
- Akbarpoor R, Khaledi N, Wang X, Samiei F. Optimization of low-energy electron beam production for superficial cancer treatments by Monte Carlo code. Journal of Cancer Research and Therapeutics. 2019;15(3):475–479. doi: 10.4103/jcrt.JCRT_203_18
- Böhlen TT, Germond JF, Desorgher L, Veres I, Bratel A, Landström E, Engwall E, Herrera FG, Ozsahin EM, Bourhis J, Bochud F, Moeckli R. Very high-energy electron therapy as light-particle alternative to transmission proton FLASH therapy — An evaluation of dosimetric performances. Radiotherapy and Oncology. 2024;194:110177. doi: 10.1016/j.radonc.2024.110177
- Böhlen TT, Germond JF, Traneus E, Bourhis J, Vozenin MC, Bailat C, Bochud F, Moeckli R. Characteristics of very high-energy electron beams for the irradiation of deep-seated targets. Medical Physics. 2021;48(7):3958–3967. doi: 10.1002/mp.14891
- Kim MM, Zou W. Ultra-high dose rate FLASH radiation therapy for cancer. Medical Physics. 2023;50(Suppl 1):58–61. doi: 10.1002/mp.16271
- Friedl AA, Prise KM, Butterworth KT, Montay-Gruel P, Favaudon V. Radiobiology of the FLASH effect. Medical Physics. 2022;49(3):1993–2013. doi: 10.1002/mp.15184
- Rosenberger P, Dagar R, Zhang W, Sousa-Castillo A, Neuhaus M, Cortes E, Maier SA, Costa-Vera C, Kling MF, Bergues B. Imaging elliptically polarized infrared near-fields on nanoparticles by strong-field dissociation of functional surface groups. The European Physical Journal D: Atomic, Molecular, Optical and Plasma Physics. 2022;76(6):109. doi: 10.1140/epjd/s10053-022-00430-6
- Ali I, Alsbou N, Ahmad S. Quantitative evaluation of dosimetric uncertainties in electron therapy by measurement and calculation using the electron Monte Carlo dose algorithm in the Eclipse treatment planning system. Journal of Applied Clinical Medical Physics. 2022;23(1): e13478. doi: 10.1002/acm2.13478
- Jeynes JCG, Wordingham F, Moran LJ, Curnow A, Harries TJ. Monte Carlo simulations of heat deposition during photothermal skin cancer therapy using nanoparticles. Biomolecules. 2019;9(8):343. doi: 10.3390/biom9080343
- Whitmore L, Mackay RI, van Herk M, Jones JK, Jones RM. Focused VHEE (very high energy electron) beams and dose delivery for radiotherapy applications. Scientific Reports. 2021;11(1):14013. doi: 10.1038/s41598-021-93276-8
- Ronga MG, Deut U, Bonfrate A, Marzi L. Very high-energy electron dose calculation using the Fermi-Eyges theory of multiple scattering and a simplified pencil beam model. Medical Physics. 2023;50(12):8009–8022. doi: 10.1002/mp.16697
- Klimpki G, Zhang Y, Fattori G, Psoroulas S, Weber DC, Lomax A, Meer D. The impact of pencil beam scanning techniques on the effectiveness and efficiency of rescanning moving targets. Physics in Medicine and Biology. 2018;63(14):145006. doi: 10.1088/1361–6560/aacd27
- Zhu J, Penfold SN. Dosimetric comparison of stopping power calibration with dual-energy CT and single-energy CT in proton therapy treatment planning. Medical Physics. 2016;43(6):2845–2854. doi: 10.1118/1.4948683
- An C, Zhang W, Dai Z, Li J, Yang X, Wang J, Nie Y. Optimizing focused very-high-energy electron beams for radiation therapy based on Monte Carlo simulation. Scientific Reports. 2024;14(1):27495. doi: 10.1038/s41598-024-79187-4
- Fan X, Niemira B. Gamma Irradiation. In: Electromagnetic Technologies in Food Science. Wiley Editors: Vicente M. Gómez-López, Rajeev Bhat. 2021. doi: 10.1002/9781119759522.ch3
- Tarim U, Gurler O, Korkmaz L. Monte Carlo simulation for the interaction characteristics of gamma-rays with several tissues and water as a tissue substitute. Radiation Effects and Defects in Solids. 2023;178(7):799–807. doi: 10.1080/10420150.2023.2184364
- Becker D, Kumar A, Adhikary A, Sevilla M. Gamma- and Ion-beam DNA Radiation Damage: Theory and Experiment. In: Radiation Chemistry: From Basic to Applications. 1st ed. Cambridge: Royal Society of Chemistry; 2020:426–455. doi: 10.1039/9781839162541-00426
- Karmaker N, Maraz K, Islam F, Haque M, Razzak M, Mollah MZI, Faruque MRI, Khan RA. Fundamental characteristics and application of radiation. GSC Advanced Research and Reviews. 2021;7(1):1–12. doi: 10.30574/GSCARR.2021.7.1.0043
- Bell BI, Vercellino J, Brodin NP, Velten C, Nanduri LSY, Nagesh PKB, Tanaka KE, Fang Y, Wang Y, Macedo R, English J, Schumacher MM, Duddempudi PK, Asp P, Koba W, Shajahan S, Liu L, Tomé WA, Yang WL, Kolesnick R, Guha C. Orthovoltage X-rays exhibit increased efficacy compared with γ-rays in preclinical irradiation. Cancer Research. 2022;82(15):2678–2691. doi: 10.1158/0008-5472.CAN‑22-0656
- Zdora M. Principles of X-ray Imaging. In: X-ray Phase-Contrast Imaging Using Near-Field Speckles. Cham: Springer; 2021:23–48.
- Khaledi N, Khan R, Gräfe JL. Historical progress of stereotactic radiation surgery. Journal of Medical Physics. 2023;48(4):312–327. doi: 10.4103/jmp.jmp_62_23
- Park HR, Jeong SS, Kim JH, Myeong HS, Park HJ, Park KH, Park K, Yoon BW, Park S, Kim JW, Chung HT, Kim DG, Paek SH. Long-term outcome of unilateral acoustic neuromas with or without hearing loss: Over 10 years and beyond after gamma knife radiosurgery. Journal of Korean Medical Science. 2023;38(40): e332. doi: 10.3346/jkms.2023.38.e332
- Nesvick CL, Graffeo CS, Brown PD, Link MJ, Stafford SL, Foote RL, Laack NN, Pollock BE. The role of biological effective dose in predicting obliteration after stereotactic radiosurgery of cerebral arteriovenous malformations. Mayo Clinic Proceedings. 2021;96(5):1157–1164. doi: 10.1016/j.mayocp.2020.09.041
- Al Saiegh F, Liu H, El Naamani K, Mouchtouris N, Chen CJ, Khanna O, Abbas R, Velagapudi L, Baldassari MP, Reyes M, Schmidt RF, Tjoumakaris S, Gooch MR, Rosenwasser RH, Shi W, Jabbour P. Frameless angiography-based gamma knife stereotactic radiosurgery for cerebral arteriovenous malformations: A proof-of-concept study. World Neurosurgery. 2022;164: e808‑e813. doi: 10.1016/j.wneu.2022.05.046
- Piwowarska-Bilska H, Kurkowska S, Birkenfeld B. Individualization of radionuclide therapies: challenges and prospects. Cancers (Basel). 2022;14(14):3418. doi: 10.3390/cancers14143418
- Barrus J, Fernando K, Addington M, Lenards N, Hunzeker A, Konieczkowski DJ. Robust VMAT treatment planning for extremity soft tissue sarcomas. Medical Dosimetry. 2023;48(4):256–260. doi: 10.1016/j.meddos.2023.06.001
- Rezaee M, Adhikary A. The effects of particle LET and fluence on the complexity and frequency of clustered DNA damage. DNA (Basel). 2024;4(1):34–51. doi: 10.3390/dna4010002
- Kumar S, Suman S, Angdisen J, Moon BH, Kallakury BVS, Datta K, Fornace AJ Jr. Effects of high-linear-energy-transfer heavy ion radiation on intestinal stem cells: implications for gut health and tumorigenesis. Cancers (Basel). 2024;16(19):3392. doi: 10.3390/cancers16193392
- Ou X, Chen X, Xu X, Xie L, Chen X, Hong Z, Bai H, Liu X, Chen Q, Li L, Yang H. Recent development in X-ray imaging technology: future and challenges. Research. 2021;2021:9892152. doi: 10.34133/2021/9892152
- Lu L, Sun M, Lu Q, Wu T, Huang B. High energy X-ray radiation sensitive scintillating materials for medical imaging, cancer diagnosis and therapy. Nano Energy. 2021;79:105437. doi: 10.1016/j.nanoen.2020.105437
- Malinowski M. Using X-ray technology to sterilize medical devices. American Journal of Biomedical Science & Research. 2021;12. doi: 10.34297/AJBSR.2021.12.001755
- Egwonor LI, Aworinde OR, Lekan OK, Osemudiamhen DA. Medical imaging: A critical review on X-ray imaging for the detection of infection. Journal of Infection Imaging. 2024. doi: 10.1007/s44174-024-00212-1
- Garnett R. A comprehensive review of dual-energy and multi-spectral computed tomography. Clinical Imaging. 2020;67:160–169. doi: 10.1016/j.clinimag.2020.07.030
- Omar A, Andreo P, Poludniowski G. A model for the energy and angular distribution of X rays emitted from an X-ray tube. Part I. Bremsstrahlung production. Medical Physics. 2020;47(10):4763–4774. doi: 10.1002/mp.14359
- Wong LWW, Shi X, Karnieli A, Lim J, Kumar S, Carbajo S, Kaminer I, Wong LJ. Free-electron crystals for enhanced X-ray radiation. Light: Science and Applications. 2024;13(1):29. doi: 10.1038/s41377-023-01363-4
- Lisovska VV, Malykhina T. Computer simulation of the angular distribution of electrons and bremsstrahlung photons in tantalum converter. Journal of Radiation Physics and Engineering. 2020;2. doi: 10.26565/2312-4334-2020-2-07
- Nakano T, Akamatsu K, Tsuda M, Tujimoto A, Hirayama R, Hiromoto T, Tamada T, Ide H, Shikazono N. Formation of clustered DNA damage in vivo upon irradiation with ionizing radiation: visualization and analysis with atomic force microscopy. Proceedings of the National Academy of Sciences USA. 2022;119(33): e2119132119. doi: 10.1073/pnas.2119132119
- Baena-Lopez L, Baonza A, Estella C, Herranz H. Editorial: Regulation and coordination of the different DNA damage responses and their role in tissue homeostasis maintenance. Frontiers in Cell and Developmental Biology. 2023;11:1175155. doi: 10.3389/fcell.2023.1175155
- Deng S, Vlatkovic T, Li M, Zhan T, Veldwijk MR, Herskind C. Targeting the DNA damage response and DNA repair pathways to enhance radiosensitivity in colorectal cancer. Cancers (Basel). 2022;14(19):4874. doi: 10.3390/cancers14194874
- Ren Y, Yang P, Li C, Wang WA, Zhang T, Li J, Li H, Dong C, Meng W, Zhou H. Ionizing radiation triggers mitophagy to enhance DNA damage in cancer cells. Cell Death Discovery. 2023;9(1):267. doi: 10.1038/s41420-023-01573-0
- Pogue BW, Zhang R, Cao X, Jia JM, Petusseau A, Bruza P, Vinogradov SA. Review of in vivo optical molecular imaging and sensing from x-ray excitation. Journal of Biomedical Optics. 2021;26(1):010902. doi: 10.1117/1.JBO.26.1.010902
- Lim CH, Lee J, Choi Y, Park JW, Kim HK. Advanced container inspection system based on dual-angle X-ray imaging method. Journal of Instrumentation. 2021;16: P08037. doi: 10.1088/1748-0221/16/08/P08037
- Flay N, Brown S, Sun W, Blumensath T, Su R. Effects of off-focal radiation on dimensional measurements in industrial cone-beam micro-focus X-ray computed tomography systems. Precision Engineering. 2020;68:93–104. doi: 10.1016/j.precisioneng.2020.08.014
- Ahmed SK, Grams MP, Locher SE, McLemore LB, Sio TT, Martenson JA. Adaptation of the Stanford technique for treatment of bulky cutaneous T-cell lymphoma of the head. Practical Radiation Oncology. 2016;6(3):183–186. doi: 10.1016/j.prro.2015.10.021
- Schüler E, Eriksson K, Hynning E, Hancock SL, Hiniker SM, Bazalova-Carter M, Wong T, Le QT, Loo BW Jr, Maxim PG. Very high-energy electron (VHEE) beams in radiation therapy: Treatment plan comparison between VHEE, VMAT, and PPBS. Medical Physics. 2017;44(6):2544–2555. doi: 10.1002/mp.12233
- Al-Shareef JM, Abousahmeen AM, Saud MAB, Al-Aqmar DM, Elfagieh M, Alwoddi BA, Adam AA, Eltayef NE, Saied FSB, Makki AM, Saleem AB. Comparison of photon versus electron for tumor bed boost radiotherapy post-breast conserving surgery. Journal of Medical Imaging and Radiation Sciences. 2023;54(3):421–428. doi: 10.1016/j.jmir.2023.05.003
- Lee VWY, Liu ACH, Cheng KW, Chiang CL, Lee VH. Dosimetric benefits of 3D-printed modulated electron bolus following lumpectomy and whole-breast radiotherapy for left breast cancer. Medical Dosimetry. 2023;48(1):37–43. doi: 10.1016/j.meddos.2022.10.001
- Hussein M, Heijmen BJM, Verellen D, Nisbet A. Automation in intensity modulated radiotherapy treatment planning: A review of recent innovations. British Journal of Radiology. 2018;91(1092):20180270. doi: 10.1259/bjr.20180270
- Roa D, Kuo J, Moyses H, Taborek P, Tajima T, Mourou G, Tamanoi F. Fiber-optic based laser wakefield accelerated electron beams and potential applications in radiotherapy cancer treatments. Photonics. 2024;9(6):403. doi: 10.3390/photonics9060403
- Lazzarini C, Grittani G, Valenta P, Zymak I, Antipenkov R, Chaulagain U, Goncalves LVN, Grenfell A, Lamač M, Lorenz S, Nevrkla M, Špaček A, Šobr V, Szuba W, Bakule P, Korn G, Bulanov SV. Ultrarelativistic electron beams accelerated by terawatt scalable kHz laser. Physics of Plasmas. 2024;31(4):043106. doi: 10.1063/5.0189051
- Ebel K, Bald I. Low-energy (5–20 eV) electron-induced single and double strand breaks in well-defined DNA sequences. Journal of Physical Chemistry Letters. 2022;13(22):4871–4876. doi: 10.1021/acs.jpclett.2c00684
- Frankl M, Macián-Juan R. Monte Carlo simulation of secondary radiation exposure from high-energy photon therapy using an anthropomorphic phantom. Radiation Protection Dosimetry. 2016;168(4):537–545. doi: 10.1093/rpd/ncv381
- Romano F, Bailat C, Jorge PG, Lerch MLF, Darafsheh A. Ultra-high dose rate dosimetry: Challenges and opportunities for FLASH radiation therapy. Medical Physics. 2022;49(7):4912–4932. doi: 10.1002/mp.15649
- Winterhalter C, Lomax A, Oxley D, Weber DC, Safai S. A study of lateral fall-off (penumbra) optimisation for pencil beam scanning (PBS) proton therapy. Physics in Medicine and Biology. 2018;63(2):025022. doi: 10.1088/1361-6560/aaa2ad
- Ohsawa D, Hiroyama Y, Kobayashi A, Kusumoto T, Kitamura H, Hojo S, Kodaira S, Konishi T. DNA strand break induction of aqueous plasmid DNA exposed to 30 MeV protons at ultra-high dose rate. Journal of Radiation Research. 2022;63(2):255–260. doi: 10.1093/jrr/rrab114
Supplementary files